Rotor and Tube Holders Included
HORIZON 6 FA
Fixed Angle Veterinary Centrifuge
Max RCF: 1,850 xg
Max RPM: 3,900 RPM
Weight: 11 lbs (5 kg)
Dimensions: 8.75 x 14 x 11.75 in (22 x 35.5 x 30 cm)
Warranty: 2 years (lifetime on rotor)
- Modeled on the world's best selling clinical laboratory centrifuge
- Simple, two-button controls
- Preset for common veterinary needs
- Designed, built, and supported in the USA
$363.09
Spin microhematocrit tubes
Spin up to 8 microhematocrit tubes with our microhematocrit adapter kit, designed to fit either 40mm or 75mm microhematocrit tubes. The adapter kit includes two (2) orange microhematocrit inserts, six (6) blue tube holders, six (6) pediatric tube adapters, and veterinary microhematocrit card reader.
Easy to Use
Spin samples with the press of just one button. The HORIZON 6 FA processes 125 mm tubes. Features a single preset runtime of 2 minutes at 370 xg.
Engineered for Safety
Protect your lab with the HORIZON 6 FA's clear, safety locking lid. The unique convection-based cooling system and brushless motor ensure years of worry-free operation.
A Great Value
Get the HORIZON 6 FA at a price that fits any budget. The unit arrives ready for use with rotor, tube holders, and 2-year warranty included in the affordable price.
The HORIZON 6 FA is a compact fixed angle veterinary centrifuge. Easy to use with simple controls designed for maximum repeatability. Small footprint for an easy fit in even the tightest labs.
- Capacity – Up to six (6) 17 mm x 125 mm tubes
- Tube Sizes – 15 mL/125 mm
- Electronic Timer – Preset to 2 minutes (+-2%)
- Brushless AC Motor – Years of operation with no routine maintenance
- Quiet Operation – A double-encased motor and the “stability mount” design provides excellent imbalance stability
- Cool Operation – Unique design of the convection based cooling system protects the samples even under the heaviest workloads
- Safety Lid Switch – Centrifuge will not run with the lid open
- Small Footprint – Compact design uses approximately 1 sq. ft. of counter space
- Automatic Lid Lock – Centrifuge will not run with the lid open and the lid cannot be opened while samples are spinning
- Positive Lid Latching – Firm mechanical latching of the lid prevents the escape of debris
- Clear Shatter Proof Lid – Allows for safe sample observation and optical calibrations
- Reinforced Guard Bowl – Provides maximum safety and durability
- Height – 8.75 in. (22 cm)
- Width – 11.75 in. (30 cm)
- Depth – 14.0 in. (35.5 cm)
- Weight – 11 lbs. (5 kg)
- Max. Acceleration Time – 10 Seconds
- Timer – Electronic, 1 to 30 Minutes (+/-2%); Preset to 2 Minutes
- Voltage requirement – 115 (+/-10) Volts
- Frequency – 60 Hz
- Protection Breaker – 4 Amp. Re-settable
- Centrifuge Motor – 1/30 H.P. Brushless AC
- Max Speed – 3,900 RPM
- Max G-Force – 1,850 xg (RCF)
Cycle 1: 2 minutes at 2,000 RPM/500 XG
Cycle 2: 10 minutes at 3,900 RPM/1,600 XG
Cycle 3: 6 minutes at 3,900 RPM/1,550 XG
All cycles can be reprogrammed to your lab's preferred settings if desired.
Manuals & Literature
Coming soon!
What is the warranty on Drucker's centrifuges?
Drucker Diagnostics provides a two (2) year parts and labor warranty (except the rotor - see question below).
What is the warranty on the rotor?
All rotors carry a lifetime warranty. Should the rotor ever fail, for as long as you own your centrifuge, Drucker Diagnostics will replace it free of charge.
Will I need to purchase accessories for the centrifuge separately?
All of our centrifuges are sold complete with the rotor and tube holders. No additional accessories are necessary though there are additional options available for purchase.
How often should my centrifuge be calibrated?
Drucker Diagnostics recommends the centrifuge be calibrated every 2 years or 5,000 cycles.
What should be used to clean and disinfect the HORIZON 6 FA?
ONLY isopropyl alcohol, soap and water, or a 10% (5,500 PPM) bleach solution should be used for cleaning and disinfection of the centrifuge and accessories. All surfaces must be dried immediately after cleaning and disinfecting.
What cleaning solutions should be avoided?
TBQ Germicidal products should not be used as they will cause damage to the centrifuge and void the warranty. The use of fully/partially halogenated hydrocarbons, ketones, esters, ethers, benzyls, ethyl benzenes, and all other chemicals not prescribed by Drucker Diagnostics should not be used as they may cause damage to the rotor chamber, rotor, tube holders, accessories, and centrifuge exterior. If this occurs the warranty will be voided.