HORIZON 6
Compact Routine Centrifuge
Max Speed: 3820 RPM / 2000 xg
Max Capacity: 6 (up to 16 x 100 mm or 10 mL)
Dimensions (W x D x H): 12 x 14 x 9 in (30.5 x 35.6 x 22.9 cm)
Weight: 12 lbs (5.4 kg)
Warranty: 2 years
- 3 cycles preset for routine lab needs
- Set and lock controls to one setting
- LED lid lighting indicates cycle status
- Designed, built, and supported in the USA
$1,470.52
Easy to Use
Use the simple, two-button interface to select a preset, or program and record your validated settings on the included label. Spin 6 tubes up to 16 x 100 mm or 10 mL in the same holder.
Easy to Monitor
Engineered for Safety
Unique Features
- Max Capacity – 6 (up to 16 x 100 mm or 10 mL)
- Factory Presets – Three (3) easily selectable preset cycles (default or custom) based on CLSI guidelines and tube manufacturer IFU
- Set and Lock Controls – The two-button interface allows users to select a cycle, or lock to a single cycle with the press of a button
- Intuitive Lid Lighting – No more forgotten samples! Lid lights flash at the end of each cycle and continue flashing until the lid is opened
- Automatic Lid Lock – Centrifuge will not run with the lid open and the lid cannot be opened while samples are spinning
- 2-Stage Horizontal Rotor – Drucker Diagnostics’ patented swing-out rotor design produces ideal horizontally separated samples
- Cool Operation – Unique design of the convection based cooling system protects the samples even under the heaviest workloads
- Brushless DC Motor – Years of operation with no routine maintenance
- Clear Shatter Proof Lid – Allows for safe sample observation and optical calibrations
- Reinforced Guard Bowl – Provides maximum safety and durability
- Quiet Operation – Enjoy a sound level of just 64 dBA
- Regulatory Approvals – Certified to US, Canadian, and European Safety regulations and standards
- Dimensions (W x D x H) – 12 x 14 x 9 in (30.5 x 35.6 x 22.9 cm)
- Weight – 12 lbs. (5.4 kg)
- Max. Acceleration Time – 20 Seconds
- Max. Deceleration Time – 20 Seconds
- Factory Setting Range – 1 to 30 Minutes (± 1%)
- Voltage – 95 to 253 VAC
- Frequency – 50/60 Hz
- Max. Power Requirements – 220 Watts
Preset 1: 10 min / 3600 RPM / 1800 xg
Preset 2: 15 min / 3300 RPM / 1500 xg
Preset 3: 5 min / 1900 RPM / 500 xg
Centrifuge Accessories
Manuals & Literature
Drucker Diagnostics Centrifuge Reference Guide
Drucker Diagnostics Centrifuge Decision Tree
HORIZON 6 Operation and Troubleshooting Guide
HORIZON 6, 6 FA, 12, and 24 Operator’s Manual (English)
HORIZON 6, 6 FA, 6 Flex, 6 Flex FA Service Manual (English)
What is the warranty on Drucker’s centrifuges?
Drucker Diagnostics provides a two (2) year parts and labor warranty.
Will I need to purchase accessories for the centrifuge separately?
All of our centrifuges are sold complete with the rotor and tube holders. No additional accessories are necessary though there are additional options available for purchase.
How often should my centrifuge be calibrated?
Drucker Diagnostics recommends the centrifuge be calibrated every 2 years or 5,000 cycles.
What should be used to clean and disinfect the Horizon 6?
ONLY isopropyl alcohol, soap and water, or a 10% (5,500 PPM) bleach solution should be used for cleaning and disinfection of the centrifuge and accessories. All surfaces must be dried immediately after cleaning and disinfecting.
What cleaning solutions should be avoided?
TBQ Germicidal products should not be used as they will cause damage to the centrifuge and void the warranty. The use of fully/partially halogenated hydrocarbons, ketones, esters, ethers, benzyls, ethyl benzenes, and all other chemicals not prescribed by Drucker Diagnostics should not be used as they may cause damage to the rotor chamber, rotor, tube holders, accessories, and centrifuge exterior. If this occurs the warranty will be voided.
Related Products
Want to learn more? Request more information
"*" indicates required fields