Executive Summary

Background: Blood sample preparation is essential for accurate diagnostic test results, which influence about 70% of medical decisions . Centrifugation is a key method for stabilizing blood samples, particularly for producing Platelet Poor Plasma (PPP), crucial for tests like coagulation profiles. This study compares the performance of two widely used centrifuges: the Drucker Diagnostics DASH Coag and the Beckman Coulter StatSpin® Express 3.

Study Overview: Sample Preparation: Blood samples from 24 volunteers were processed using both centrifuges. The resulting PPP was analyzed for platelet counts using a Sysmex XN-1000 Hematology Analyzer.

Performance: Both centrifuges produced consistent PPP with platelet counts under 10 x103 platelets/μL. Statistical analysis, including ANOVA (Analysis of Variance), confirmed no significant difference between the two centrifuges.

Key Findings: PPP Production: The DASH Coag and StatSpin® Express 3 are equivalent in producing high-quality PPP.

Design Differences: The DASH Coag has a larger capacity, while the StatSpin® Express 3 is more compact. Both are user-friendly and compliant with safety standards.

Conclusion: The Drucker Diagnostics DASH Coag and the Beckman Coulter StatSpin® Express 3 are both highly effective in preparing Platelet Poor Plasma, with no significant difference in their performance as demonstrated by this study. Laboratories can confidently choose either centrifuge based on specific needs such as capacity, size, and operational preferences, without compromising on the quality of PPP produced.

Nuo Aurix custom centrifuge